Calibration procedure for Weighing Balance.
|
A |
weighing balance is essential to ensure
accurate measurements. Here’s a step-by-step procedure for calibration:
1.Repeatability
Repeatability
measures the variation in multiple measurements taken under the same
environmental conditions.
Perform this test
for both full load and half load of the balance.
Follow these steps:
Press the Zero button
on the balance.
Place the weight
(either half load or full load) on the weighing pan.
Note down the
reading.
Remove the weight
without zeroing the balance.
Repeat the process
at least 10 times.
Calculate the standard
deviation.
2.Linearity:
Linearity assesses
the deviation from the nominal value of weight.
Ideally, the graph
between standard weight and indicated reading should be a straight line.
Take readings for
different weights (including no load and full load):
Press the Zero button
on the balance.
Place the minimum
weight on the weighing pan and note the reading.
Continue with
increasing and decreasing order of weights.
Repeat this step
twice.
Calculate the
average for each weight reading.
Linearity =
Standard Reading - Weighing Balance (UUC) Reading (UUC is Unit Under
Calibration).
3.Eccentricity:
Ideally, the
weighing object should be placed exactly at the center of the weighing pan.
However, this is
not always possible.
Follow these steps:
Place a weight in
the center of the weighing pan.
Observe the
displayed value.
Repeat this
procedure nine times for the standard weight and record the readings.

Comments
Post a Comment